N9030B PXA Signal Analyzer Service Guide 553
Post-Repair Procedures
Post-Repair Procedures
Table 17-1
Post-Repair Testing Requirements
Assembly
Adjustments to Perform
(in the order listed)
a
Performance Verifications to Perform After
Adjustments
A1A2
Front Panel
Interface
None
Residual Responses
Displayed Average Noise Level
A1A3
Liquid Crystal
Display
None
Residual Responses
Displayed Average Noise Level
A2
Analog IF Assembly
IF Flatness Adjust (Option B25, B40, or B1X if
Option MPB is also present)
Power Bandwidth Accuracy
Resolution Bandwidth Switching Uncertainty
Residual Responses
Displayed Average Noise Level
IF Frequency Response (Option B25, B40, B1X)
Absolute Amplitude Accuracy
Display Scale Fidelity
A3
Digital IF Assembly
FPGA Synchronization
b
Power Bandwidth Accuracy
Resolution Bandwidth Switching Uncertainty
Residual Responses
Displayed Average Noise Level
IF Frequency Response (Option B25, B40, B1X)
Absolute Amplitude Accuracy
Display Scale Fidelity
A4
CPU Assembly
BIOS Settings Verification
Set Instrument Time and Date
c
Residual Responses
Displayed Average Noise Level
A4BT1
Battery
BIOS Settings Verification
Set Instrument Time and Date
Residual Responses
Displayed Average Noise Level
A5
Disk Drive
d
Accept End User License Agreement (EULA)
Instrument Software Update
e
All Adjustments
f
Residual Responses
Displayed Average Noise Level
Frequency Response (All Tests)
A6
Power Supply
None
Residual Responses
Displayed Average Noise Level
Spurious Responses
A7
Midplane board
Program Instrument Model and Serial
Number
Residual Responses
Displayed Average Noise Level
A8
Motherboard
None
Run all tests since instrument was disassembled.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...