N9030B PXA Signal Analyzer Service Guide 243
Analog/Digital IF Troubleshooting
A3 Digital IF Troubleshooting
Verifying the 250 MHz IF Input Power (Option B40 only)
1.
Perform an instrument shutdown.
2.
Remove the cover of the PXA Signal Analyzer. Refer to
“Assembly Replacement Procedures”, on page 429
in this service guide.
3.
Turn the instrument over so that the bottom side of the analyzer is facing
up.
4.
Turn on the PXA Signal Analyzer and wait for the instrument to complete
the boot up process.
5.
Press
Mode
,
IQ Analyzer (Basic)
,
Mode Setup
,
IF Path
,
40 MHz
.
6.
Press
System
,
Alignments
,
Auto Align
,
Off
.
7.
Press
Input/Output
,
RF calibrator
,
50 MHz
.
8.
Verify the 50 MHz signal is at
−
25 dBm by pressing
FREQ
,
50 MHz
,
SPAN
,
1 MHz
,
Peak Search
on the analyzer. The maker readout should be 50 MHz
at
−
25 dBm
±
3 dB. If this reference signal is measuring incorrectly, see
Chapter 4, “RF Section Troubleshooting (RF/Microwave Analyzers)”, on
in this service guide.
9.
To continue verifying press
Meas
,
IQ Waveform
. Verify the input
attenuator on the PXA is set to 10 dB. Look near the top of the display near
the center and verify
Atten: 10 dB
is visible. If needed change the input
attenuator by pressing
AMPTD
,
Attenuation
,
10 dB
on the analyzer
10.
Referring to
, carefully disconnect W26 or W40 (if analyzer is
also equipped with Opt B1X) from A3J15.
Perform this procedure only if the analyzer is equipped with Option B40,
Analysis Bandwidth, 40 MHz.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...