226 N9030B PXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Assembly Description
A2 Analog IF Assembly Description
When the 10 MHz or 25 MHz IF Path is selected, or when any swept
measurement is made, the analyzer's RF input signal is down converted to a
322.5 MHz intermediate frequency in the A13 RF Front End Assembly. This
322.5 MHz signal is routed through the IF MUX on the A15 Front End Control
Assembly and then input to the A2 Analog I.F. Assembly.
The Analog I.F. down converts the 322.5 MHz signal to a final 22.5 MHz
intermediate frequency. In order to obtain optimal dynamic range and
minimize unwanted spurious signals, this signal is filtered, amplified and
attenuated throughout this assembly. The final 22.5 MHz signal goes through a
series of variable band pass filters each having a specific bandwidth. Three
things determine the signal path and which filters are chosen:
— Instrument resolution bandwidth (RBW) setting
— Application used
— Mode applied
This assembly contains one of the two burst carrier trigger (BCT) circuits for
performing mobile communication measurements. The BCT on the A2
assembly is relatively narrowband. The other BCT circuit is part of the IF MUX
on the A15 Front End Control Assembly and has a much wider bandwidth.
The output of the A2 assembly is a filtered 22.5 MHz signal that goes to one of
the two ADCs on the A3 Digital I.F. Assembly.
Inputs to the A2
322.5 MHz signal (from A15 Front End Control Assembly)
LO Input (300 MHz signal from A16 Reference Assembly)
22.5 MHz Cal Comb Signal (from the A3 Digital IF Assembly)
Outputs from the A2
22.5 MHz signal (to A3 Digital IF Assembly)
RF Trigger Detector (to A3 Digital IF Assembly)
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...