98 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–312
PUD memory lost
Protected user data saved by the *PUD command has been lost.
–313
Calibration memory
lost
The nonvolatile calibration data used by the *CAL? command has been lost.
–314
Save/recall memory
lost
The nonvolatile data saved by the *SAV? command has been lost.
–315
Configuration memory
lost
The nonvolatile configuration data saved by the instrument has been lost.
–320
Storage fault;
A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
–321
Out of memory
An internal operation needed more memory than was available. Report this
error to the nearest Keysight Technologies sales or service office.
–330
Self-test failed
A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office.
–340
Calibration failed
The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
–340
Calibration failed;
<failure msg> <port>
The calibration for one of the I-Q ports did not succeed. The information in
the “failure msg” field can be used to troubleshoot this problem. Contact
Keysight technical support if necessary.
–350
Queue overflow
An error occurred that did not get put in the error queue because the
queue was full.
–360
Communication error
There was a problem with instrument remote communications. The exact
problem cannot be specifically identified.
–360
Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
The Keysight Smart Noise Source connected to the PXA has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or
another hardware fault exists. Check SNS on another instrument, NFA and
ESA are also SNS compatible instruments. Check the device is not an
Keysight power sensor which uses the same cable interface.
–361
Parity error in
program message
A parity bit was not correct when the data was received. For example, on a
parallel port.
–362
Framing error in
program message
A stop bit was not detected when data was received. For example, on a
remote bus port.
–363
Input buffer overrun
A software or hardware input buffer on a port overflowed with data
because of improper or nonexistent pacing.
–365
Time out error
There was a time-out problem in the instrument. The exact problem cannot
be specifically identified.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...