N9030B PXA Signal Analyzer Service Guide 107
Instrument Messages
Event Messages
–221
Settings conflict; Span Zoom
is not available in Zero
Span
Span Zoom does not work with a time domain x-axis. You
must select a span greater then 0 Hz.
–221
Settings conflict; Span Zoom
not available when Frequency
Scale Type = Log
The Span Zoom feature cannot be used when the X-axis
scale is logarithmic frequency.
–221
Settings conflict; Step keys
are not available to modify
this function
You should select a specific value for this function. So using
the Up/Down step keys to scroll through values is not
allowed.
–221
Settings conflict; Sweep
Setup only available in
swept measurements
The current measurement uses FFT mode and so does not
use the Sweep Setup menu
–221
Settings conflict; Sweep
Time cannot be auto-coupled
in FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
–221
Settings conflict; Sweep
Time cannot be auto-coupled
while in Zero Span
You cannot send the remote command to set the sweep time
to auto while you are in zero span.
–221
Settings conflict; Sweep
Time cannot be set while in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. It
cannot be manually controlled.
–221
Settings conflict; Swept IF
Gain High not available when
FFT IF Gain = Manual Low
When FFT IF Gain is manually set to Low, you cannot set the
Swept IF Gain to High because that would make the
Reference Level couplings wrong in swept mode.
–221
Settings conflict; Swept LO
not available when freq mode
is Fixed
SCPI only message. The LO Mode cannot be set to Swept
when the freq mode is set to fixed. Change the freq mode
away from fixed, or perform the measurement at several
fixed frequencies.
–221
Settings conflict; Swept
Type=Swept is not available
while in Gated FFT
If you have selected gated FFT then you are using the FFT
sweep type and you cannot select the swept type of
sweeping.
–221
Settings conflict; System
Display Settings, Annotation
is Off
This is an override that turns off many of the annotations.
This is available as a security feature.
–221
Settings conflict; T hot
must be greater than T cold
The Tcold value set under Meas Setup/ENR/Tcold, needs to
be lower than the Thot value currently being set. Tcold is
often taken as the ambient temperature of the noise source.
If using an SNS the Tcold value may be read automatically
before every sweep.
–221
Settings Conflict; The
parameter cannot be changed
in FAST mode
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...