266 N9030B PXA Signal Analyzer Service Guide
LO Synthesizer/Reference Troubleshooting
A16 Reference Assembly Description
A16 Reference Assembly Description
Purpose
The A16 Reference Assembly provides the fundamental reference signals from
which all instrument local oscillator and synchronous timing signals are
derived. The reference board also uses these same signals to generate CW and
modulated RF calibration signals and a 10 MHz time base signal for the
instrument.
The Reference assembly must be replaced as a single assembly, although it is
constructed of two boards, the A16 Reference Main board and A16A1
Reference Daughter board. These boards are listed individually in the Show
Hardware screen, allowing identification of the individual board part numbers
comprising the complete assembly.
The reference signals and local oscillators are phase locked to a timebase in
one of two ways:
1.
The internal 10 MHz reference located on the A16 Main board.
2.
An externally provided reference such as a house standard or the reference
from another instrument or a base transceiver system. The external
reference can be any RF signal in the range of 1 MHz to 50 MHz, -5 dBm to
+10 dBm.
If an external reference signal is within 200 Hz of the specified external
reference frequency and
Freq Ref In
is set to
Sense
, the analyzer will
automatically switch from SENSE:INT to SENSE:EXT.
The phase noise performance of the reference signals establishes the limit on
attainable phase noise performance of the LO signals derived from them.
Therefore, excellent reference signals phase noise is required to achieve
excellent LO signal phase noise.
Status monitoring is comprehensive. Every PLL is monitored for phase lock and
every ALC is monitored for correct leveling. A set of LED's which indicate
whether the 100 MHz VCXO PLL and the 2400 MHz CRO PLL is phase locked
or unlocked. Each oscillator (10 MHz, 100 MHz, 2400 MHz) is monitored for the
presence of RF power (dead/alive status).
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...