138 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
309
4
User Cal; Cal invalidated
E,W
The existing user cal has been invalidated
because of one of the following reasons:
Frequency: Setting the frequency outside
the current valid user cal set (for
example: If the current sweep range is 2
to 3GHz, then setting the start frequency
to 1.9 GHz will invalidate the current user
cal. Other frequency changes that will
invalidate the user cal are:
• If the cal was performed at a
fixed frequency and you change
this frequency.
• If you are in “Freq List” mode
and you change it to extend
beyond the current user cal
range. In this case you will see
an error message.
DUT Type: If the DUT Type parameter
changes, causing the measurement
frequencies to be pushed outside the
current cal.
Attenuation: If an attenuation setting is
selected but has not been calibrated.
Preamp: If set to condition different from
current cal settings, for example: if
calibrated with the preamp on, turning it
off will invalidate the cal.
Points: Changing the number of
measured frequency points can make the
stored preselector offsets become
inaccurate and hence invalidate the
calibration. This occurs when the
following conditions exist:
• A successful calibration has
been performed.
• Some measured freq points are
> 3.6 GHz.
• The new points > 3.6 GHz are
located more than 50 MHz
away from the current
calibration points.
Err#
Status
Register
Bit
Message
Error or
Warning
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...