Instrument Messages
Event Messages
–124
Too many digits
The mantissa of a decimal-numeric contained more than 255
digits, excluding leading zeros.
–128
Numeric data not allowed
A legal numeric data element was found, but that is not a valid
element at this position in the command.
–130
Suffix error
A problem was found in a suffix (units). The exact problem
cannot be specifically identified.
–131
Invalid suffix
There is a syntax problem with the suffix. You need to use the
suffix (units) that are allowed by this command.
–134
Suffix too long
The suffix contained more than twelve characters.
–138
Suffix not allowed
A suffix was found after a numeric element that does not allow
suffixes (units).
–140
Character data error
A problem was found with a character data element. The exact
problem cannot be specifically identified.
–141
Invalid character data
Either the character data element contains an invalid character
or the element itself is not valid for this command.
–144
Character data too long
The character data element contains more than twelve
characters.
–148
Character data not
allowed
A character data element that you sent is valid, but it is not
allowed in this point in the parsing.
–150
String data error
A problem was found with a string data element. The exact
problem cannot be specifically identified.
–151
Invalid string data
A string type of data element was expected, but it is invalid for
some reason. For example, an END message was received
before the terminal quote character.
–158
String data not allowed
A string data element that you sent is valid, but it is not allowed
at this point in the parsing.
–160
Block data error
A problem was found with a block data element. The exact
problem cannot be specifically identified.
–161
Invalid block data
A block data element was expected, but it was invalid. For
example, an END message was received before the end length
was satisfied.
–168
Block data not allowed
A legal block data element was found, but it is not allowed at this
point in the parsing.
–170
Expression error
A problem was found with an expression data element. The exact
problem cannot be specifically identified.
–171
Invalid expression
An expression data element is not valid. For example, there may
be unmatched parentheses or an illegal character.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...