208 N9030B PXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Power Supply Verification
Looking at the test board, verify power supply green LED's DS5, DS6, DS7 and
DS8 located under the J2 connector are turned on. Additionally red LED's DS9
and DS10 should also be on. The power supply voltages are listed in the table
below.
If any of the power supply LED's are not turned on, see
and Initialization Troubleshooting”
.
Input Attenuator A Control Logic Verification (Option 503, 508, 513,
526)
Press
AMPTD
,
Attenuation 0 dB
on the analyzer.
Attenuation LED's DS15-20 on the test board should be off.
When the input attenuation is changed from 0 dB to 2 dB, the 2 dB Step LED
DS15 should illuminate. When the input attenuation is changed from 2 dB to
4 dB, DS15 and DS16 should be illuminated as per
If this is incorrect, the most probable cause is the A15, Front End Control
assembly.
The analyzer default setting for RF Coupling is AC mode. Verify the AC Select
LED DS14 is illuminated at this time. Press
Input/Output
,
RF Input
,
RF
Coupling
,
AC/DC
to DC on the analyzer. DS14 should turn off and DS11, DC
Select LED should illuminate. Switch the RF Coupling back to AC. If the LED's
do not illuminate correctly, the most probable cause is the A15 Front End
Control Assembly.
Table 6-4
Power Supply Voltages
LED
Power Supply
DS5
+15 VDC
DS6
+9.0 VDC
DS7
+5.2 VDC
DS8
-15 VDC
DS9
+25 VDC
DS10
+25 VDC
Table 6-5
Input Attenuator A LED’s
Attenuator Setting (dB)
DS15
DS16
0
OFF
OFF
2
ON
OFF
4
ON
ON
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...