90 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Advisory Messages
Frequency Hopping enabled,
waiting for valid burst
The demodulated burst type has not been found in the originally
demodulated slot location within the frame.
Frequency menu has changed to
reflect frequency context
switch
The frequency context parameter has been changed either by the user or
the system. The frequency menu will now contain the frequencies for the
new context. No action required.
Gate required for valid
results
High-pass and Low-pass
filters set to OFF
Turning on any band pass filter will turn off high-pass and low-pass filters.
High-pass filter set to OFF
Turning on any band pass filter will turn off high-pass filters.
IF Fixed freq should be
greater than LO Stop freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The IF fixed freq should be greater than the LO Stop
freq for a USB upconverter swept LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
IF Fixed freq should be
greater than RF Stop freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The IF fixed freq should be greater than the RF Stop freq
for an upconverter swept LO setup. Use the graph icon on the DUT setup
form to clarify the setup required.
IF Start freq should be
greater than LO Fixed freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The IF start freq should be greater than the LO fixed freq
for an USB upconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
IF Start freq should be
greater than RF Start freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The IF start freq should be greater than the RF Start
freq for an upconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
Input is internal
The instrument’s input is set to internal (the internal amplitude reference
signal). So any signals connected to the front/rear panel inputs cannot be
measured.
LO Fixed freq should be
greater than IF Stop freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO fixed freq should be greater than the IF Stop
freq for an LSB upconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
Message
Description/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...