Instrument Messages
Event Messages
–221
Settings conflict;
Destination trace for Trace
Math cannot be a trace
operand
The resulting trace data (from doing a trace math function)
cannot be put into the any of the traces that are being used
by the math operation.
–221
Settings conflict; Downconv
only available when DUT is
Amplifier
SCPI only message. The System Downconverter can only be
set to ‘On’ when the DUT type is amplifier. Change DUT type
to Amplifier if the System Downconverter is required.
–221
Settings conflict; EDGE EVM
only supports EDGE TCH burst
type.
–221
Settings conflict;
Electronic attenuator is
disabled
You are using the mechanical attenuator, and have not
enabled the electronic attenuator. You cannot set the value
of the electronic attenuator because it automatically
sets/changes when enabled.
–221
Settings conflict;
Electronic attenuator is not
available above 3.6 GHz
The maximum frequency of the electronic attenuator is 3.6
GHz. This is because of switching capacitance.
–221
Settings conflict;
Electronic attenuator
unavailable in current state
–221
Settings conflict;
Electronic attenuator
unavailable with Preamp on
The internal preamp is on. Electronic attenuator cannot be
used while you are using the internal preamp.
–221
Settings Conflict; FAST
method can only be used
while Radio Std is W-CDMA
–221
Settings conflict; Feature
not available in this View
Some functionality is available in one View, but not in
another. (See the Views under the View/Display key.) This
error occurs if you send a SCPI command or push a gray-out
key that is not available in the current selected View.
–221
Settings conflict; Feature
not supported for selected
source
You have asked for a feature that the selected source is not
capable of.
–221
Settings conflict; Feature
not supported for this
measurement.
Some functionality is available in one measurement, but not
in another. (See the measurements under the Meas key.)
This error occurs if you send a SCPI command or push a
gray-out key that is not available in the current selected
measurement.
–221
Settings conflict; Feature
not supported for this model
number
This functionality is not a part of the instrument you are
using but may be found in other models in the X-Series.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...