374 N9030B PXA Signal Analyzer Service Guide
Service and Diagnostics Menus
Service Key Descriptions
This is exactly the same as the customer accessible Align Now, RF
with one exception: the alignment can be invoked even if a previous
Align Now, All has not been executed (this key is not grayed-out if
using default alignment data). See description of Align Now, RF in the
User's Guide.
Current System
Gain
Immediately executes an alignment of the Current System Gain, for
the purpose of improving small amplitude variations that occur as
resolution bandwidth is switched. This alignment is done by
measuring the response of the current system state configuration to
the 50 MHz amplitude reference signal. All subsequent
measurements are then compensated appropriately for absolute
amplitude accuracy. The instrument will stop any measurement
currently underway, perform the alignment, then restart the
measurement from the beginning.
The Advisory Event “Alignment complete” is displayed when the
alignment is complete.
Align Current System Gain can be interrupted by pressing the Cancel
(ESC) front-panel key. When this occurs, no new Current System
Gain alignment data will be employed.
LO Phase Noise Adjustment
Immediately executes an optimization for LO Phase Noise. When the
optimization is finished the value is stored in non-volatile memory.
The instrument will stop any measurement currently underway,
perform the alignment, and then restart the measurement from the
beginning (similar to pressing the
Restart
key).
Nulling
The hardware is capable of nulling (cancelling) the LO Feedthru,
intermodulation products, or neither. It is physically not capable of
nulling both at the same time.
LO Feed thru Nulling
Control is available to set the state of LO Feedthru nulling ON or OFF.
In the ON state, the software will determine when it is appropriate to
null the LO Feedthru. LO nulling is only available in low band.
Intermod Nulling
Control is available to set the state of Intermod Nulling On or OFF. In
the ON state, the software will determine when it is appropriate to
null the intermodulation products. IM nulling is only available in low
band.
Diagnostics
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...