N9030B PXA Signal Analyzer Service Guide 91
Instrument Messages
Advisory Messages
LO Fixed freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO fixed freq should be greater than the RF Stop
freq’s for an LSB or DSB (for DSB measurements the setup uses LSB
values) downconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
LO Start freq should be
greater than IF Fixed freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO Start freq should be greater than the IF fixed
freq for an LSB upconverter swept LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
LO Start freq should be
greater than RF Start freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO start freq should be greater than the RF Start
freq’s for an LSB downconverter swept LO setup. Use the graph icon on
the DUT setup form to clarify the setup required.
LO Stop freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO Stop freq should be greater than the RF Stop
freq’s for a DSB (for DSB measurements the setup uses LSB values)
downconverter swept LO setup. Use the graph icon on the DUT setup form
to clarify the setup required.
Low-pass filter set to OFF
Turning on any band pass filter will turn off low-pass filters.
No spurs have been found
You has started a measurement in examine meas type in single or
continual sweep mode, or full meas type in single sweep mode, but no
spurs were found.
Preparing Calculation…
Preselector is centered
The preselector has been successfully centered
Preselector not used in this
frequency range.
You cannot center or adjust the preselector because it is not used at all at
the current marker frequency or between the current start and stop
frequencies
Probe connected, cal data is
being reapplied; <port>;
<probe>
A probe has been connected, calibration data is being reapplied
Probe connected, no probe
cal; using cable cal data;
<port>; <probe>
A probe has been connected and no probe calibration data is available.
The latest cable calibration data will be used
Probe disconnected,
reverting to cable
calibration data; <port>
A probe has been disconnected, calibration data reverting to the last cable
calibration data
Message
Description/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...