N9030B PXA Signal Analyzer Service Guide 55
Boot Up and Initialization Troubleshooting
PXA Instrument Boot Up Process
4.
The instrument fans should start running. The fans are mounted on the
left-hand side of the analyzer and draw air into the instrument to cool the
internal circuitry. If a fan is not running refer to the
section in this chapter.
5.
The Keysight Technologies splash screen is displayed in white font on a
dark background for ~5-10 seconds after the analyzer is turned on. If the
Keysight Technologies logo is not displayed refer to the
section in this chapter. If the instrument hangs
at the Keysight Technologies splash screen refer to the
at the Keysight Splash Screen”
section in this chapter.
6.
Verify text is displayed on screen where the user has the option of booting
Windows XP or running the Instrument Recovery System. The default
selection is to boot Windows XP. If a recovery is required, press the Down
Arrow key on the front panel of the analyzer within 5 seconds to highlight
“Keysight Recovery System” and press the Enter key on the analyzer,
otherwise Windows XP will begin to boot. If the Windows XP boot screen
is not displayed within a few seconds refer to the
Completely Load or Run the Operating System”
section in this chapter.
7.
If the recovery system is not selected the Windows XP Professional
operating system will begin to boot up. This will take ~20-30 seconds.
8.
The Keysight Technologies logo is displayed in white font on a blue
background while Windows finishes loading user preferences. This can
take up to 4 minutes. If this does not occur refer to the
Completely Load or Run the Operating System”
section in this chapter.
9.
By default, the initialization process of the PXA Spectrum Analyzer
application begins loading. The N9030B PXA Signal Analyzer screen
appears. This screen remains for slightly over 1 minute. If any of the
initializing processes do not complete, refer to the
section in this chapter.
10.
While the application software is loading the instrument will perform a
number of internal alignments before the analyzer begins to sweep. A
screen with a black background appears and the alignment progress is
shown in a yellow box. The number of alignments is dependent on which
hardware options are present. If an alignment fails, refer to the
section in this chapter.
If a recovery was selected follow the on-screen instructions and perform a
system recovery. Additional information about performing a system
recovery can be found in
Chapter 9, “CPU/Disk Drive Troubleshooting”, on
in this manual.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...