N9030B PXA Signal Analyzer Service Guide 95
Instrument Messages
Event Messages
–800, Operation Complete Event
–700, Request Control Event
–600, User Request Event
–500, Power on Event
–400 to –499, Query Errors
Err#
Message
Verbose/Correction Information
–800
Operation complete
The instrument has completed all selected pending operations in
accordance with the IEEE 488.2, 12.5.2 synchronization protocol.
Err#
Message
Verbose/Correction Information
–700
Request control
The instrument requested to become the active IEEE 4881
controller-in-charge.
Err#
Message
Verbose/Correction Information
–600
User request
The instrument has detected the activation of a user request local
control.
Err#
Message
Verbose/Correction Information
–500
Power on
The instrument has detected an off to on transition in its power
supply.
Err#
Message
Verbose/Correction Information
–400
Query Error
There was a problem with a query command. The exact problem
cannot be specifically identified.
–410
Query INTERRUPTED
Some condition caused an INTERRUPTED query to occur. For
example, a query was followed by DAB or GET before a response
was completely sent.
–420
Query UNTERMINATED
Some condition caused an UNTERMINATED query to occur. For
example, the device was addressed to talk and an incomplete
program message was received.
–430
Query DEADLOCKED
Some condition caused a DEADLOCKED query to occur. For
example, both the input buffer and the output buffer are full and
the analyzer cannot continue. The analyzer automatically discards
output to correct the deadlock.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...