N9030B PXA Signal Analyzer Service Guide 159
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
A9 Input Attenuator and A10 Input Attenuator Verification
Calibrator Switch Test
On the A16 Reference assembly, disconnect semi rigid cable W3 from A16J701
and measure A16J701 with a spectrum analyzer. Expected signal is 50 MHz at
−
25 dBm
±
0.5 dB. If signal level is incorrect, suspect A16 Reference assembly
is faulty, or the 50 MHz calibrator amplitude requires adjustment using the
field calibration software.
To verify calibrator switch operation, connect external signal source set to 50
MHz and
−
25 dBm to the RF input connector of analyzer under test. Press
Input/Output
,
RF Calibrator
,
Off
. If the signal level at the attenuator output is
now correct, suspect the A9 Attenuator cal switch or a faulty control signal
from the A15 Front End Controller assembly.
Attenuator Check
Set the signal source connected to the analyzer input port to 0 dBm. On the
analyzer, press
Amplitude
,
Attenuation
, and change input attenuation to 0 dB.
The measuring spectrum analyzer connected to A10 Attenuator output port
should indicate a 0 dBm level.
Change the input attenuation on the analyzer under test to 2 dB. See the chart
below for expected measurement values at the A10 Attenuator output port.
Analyzer Under Test
Attenuator Setting (dB)
Power at A10 Output
Port
Input Attenuator Being
Tested
0
0 dBm (reference)
Both set to through path
2
−
2
A9
4
−
4
A9
6
−6
A10
8
−
8
A9 set to 2 dB,
A10 set to 6 dB
10
−
10
A10
20
−
20
A10
30
−
30
A10
40
−
40
A10
50
−
50
A10
60
−
60
A10
70
−70
A9 set to 4 dB,
A10 set to 66 dB
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...