104 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–221
Settings conflict; Log Scale
Type is not available with
Demod View
The logarithmic x-axis scales are not available when you
have the demod view turned on.
–221
Settings conflict; Log Scale
Type is only available in
swept measurement
Logarithmic scaling can be used when making a swept SA
measurement. It is not available in the SA measurement
when you are using FFT sweeps.
–221
Settings conflict; Marker 1
Trace Update=off turns off
Signal Track
Signal Track not available unless the trace containing Marker
1 is updating
–221
Settings conflict; Marker
cannot be relative to itself
A marker must be set relative to another marker, not to
itself.
–221
Settings conflict; Marker
Count is not available when
Gate on
The marker count function cannot be used while you have
gating turned on.
–221
Settings conflict; Marker
Function is not available
for a Fixed marker
If a Marker is a Fixed type marker, the marker's value does
not change from when it first became fixed. You cannot turn
on or change a Marker Function because there is no ongoing
measurement data to use for the marker function
calculation.
–221
Settings conflict; Marker
type must be delta
Mkr
Δ
->Span and Mkr
Δ
->CF require that the selected
marker be a delta marker.
–221
Settings conflict; Marker->
function is not available in
zero span
Most of the “Marker To” functions are not available if you are
in zero span (span = 0 Hz, or time domain). So you cannot
send the commands for these functions.
–221
Settings conflict; Mask
unavailable for current
Span. Increase to display
mask.
The current span setting is either narrower than the mask
width or so wide that there are too few display points to
allow the mask to be drawn. Increase or decrease the span
to display the mask.
–221
Settings conflict; Meas Type
was changed to Examine for
Exp Avg Mode.
Average Mode has been changed to Exponential. Full Meas
Type is not available for Exponential Average Mode therefore
Meas Type has been changed to Examine.
–221
Settings conflict; Meas Type
was changed to Full for
Repeat Avg Mode
Average Mode has been changed to Repeat. Examine Meas
Type is not available for Repeat Average Mode therefore
Meas Type has been changed to Full.
–221
Settings Conflict;
MinPts/RBW limit not met
–221
Settings conflict; Mkr -> CF
is not available when the
x-axis is time domain
The marker to center frequency functionality does not work
when the x-axis is in the time domain.
–221
Settings conflict; MS gain
is not available in this
Mode
Mobile Station gain correction is not available in some
Modes, or in some measurements (for example, the SA
measurement).
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...