106 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–221
Settings conflict; Param
only available when
Frequency Mode is Swept
SCPI only message. This parameter is only available when
the Freq mode is set to Swept. Change the Freq Mode to
Swept
–221
Settings conflict; Param
only available when valid
cal data exists
SCPI only message. The ‘Apply Calibration’ parameter is only
available when the stored cal data matches the current
setup. Perform a fresh ‘Calibrate Now’ or change setup such
that current cal data is valid.
–221
Settings conflict; Preamp
gain is not available in
this Mode
Preamp gain correction is not available in some Modes or
Measurements
–221
Settings conflict; Preamp
unavailable with electronic
attenuator on
The electronic attenuator is on. Internal preamp cannot be
used while you are using the electronic attenuator.
–221
Settings conflict; Reference
marker must be in same
window
A delta marker and its reference must be in the same
window. This error occurs when you try to turn on a delta
marker who’s reference is in a different window.
–221
Settings conflict; Relative
Trigger needs hardware
support for this meas
To do Relative Triggering in this measurement requires
optional hardware that is not present in this analyzer.
–221
Settings conflict; Scale
Type = Lin is not available
when Normalize is on
Only the Log amplitude scale is available in Normalize, since
the results are always presented as a dB ratio.
–221
Settings conflict; Settings
conflict; Pre-trigger is
insufficient for demod.
Decrease Trig Delay.
–221
Settings conflict; Signal
Track is not available when
Freq Scale=Log
The signal tracking feature cannot be used when you have
selected a log scale for the frequency axis.
–221
Settings conflict; Signal
Track is not available with
Continuous Peak
The signal tracking feature cannot be used while you are also
using the continuous peak function.
–221
Settings conflict; Signal
Track is only available in
Swept SA measurement
The signal track functionality can be used when making a
swept SA measurement. It is not available in the SA
measurement when you are using FFT sweeps.
–221
Settings conflict; Signal
Track is turned off when
Zero Span is selected
Signal Track is not available when you have selected Zero
Span. So if Zero Span is entered while in Signal Track is On,
Signal Track is turned off.
–221
Settings Conflict; Span
limited to XXX
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...