110 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–200 to –299, Execution Errors
For –221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
–21x – Trigger errors
–22x – Parameter error
–23x – Data corrupt or stale (invalid data)
–24x – Hardware error
–25x – Mass storage error
–26x – Expression data error
–27x – Macro error
–28x – Program error (a downloaded program-related execution error)
–29x – Memory use error
Err#
Message
Verbose/Correction Information
–200
All ranges are off. Turn
on at least a range
There are no range turn on in scan table. You need to turn on at
least a range to initiate a scan.
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom.
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom.
–200
Execution Error
A program execution error has occurred. The exact problem
cannot be specifically identified.
–200
Execution error; Carrier
frequency outside device’s
transmit band
The entered channel/carrier frequency is not within the range of
your current mode setup selection of standard and device.
–200
Execution error; Invalid
GSM burst timing
A GSM-like burst was acquired, but its timing is not valid. Ensure
the correct Burst Type has been selected.
–200
Execution error; Invalid
IP address
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
–200
Execution error; Invalid
Marker Trace.
Cannot place markers on the reference trace, because the
reference trace is currently turned off or has no data.
–200
Execution Error; No peak
found.
No signal peak was found within the defined parameters of the
search.
–200
Execution error; No ranges
are defined. Activate a
range.
There are no active ranges in the range table. You will need to
activate at least one range.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...