116 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–250
Mass storage error; Can
only load an Antenna Unit
into Correction 1
The only Correction register that supports Antenna Units is
number 1. You have attempted to load an Ampcor file which
contains antenna units into another register
–250
Mass storage error; Cannot
make
The directory or file cannot be created.
–250
Mass storage error;
Different Antenna Unit
already in use
Attempt to import Corrections file with Antenna Unit that differs
from an in-use correction.
–250
Mass storage error;
Directory not found
The system cannot find the path specified.
–250
Mass storage error; Failed
to Load trace. Bad file
format.
The load trace operation could not be completed, as the input file
was not in the expected format. You can only load traces that were
previously saved using the 'Save Trace' feature.
–250
Mass storage error; File
<filename> wrong type
Attempt to import a data file that is not the proper type for this
operation.
–250
Mass storage error; File
<filename> and instrument
version mismatch
While opening a file, there was a mismatch between file version or
model number with instrument version or model number. The
import still tried to load as much as possible, but you should
check it closely.
–250
Mass storage error; File
contains incorrect data
for this operation
There is a mismatch between the file data type of the file specified
and the destination indicated. For example, a correction set
cannot be loaded/imported into a limit line.
–250
Mass storage error; File
empty
Cannot save trace because it contains no data. Check that the
trace is turned on and contains some valid data.
–250
Mass storage error;
Invalid register number
for *SAV or *RCL Mass
Storage error
You have used the *SAV command to save a state to a non
existent state register.
Or
You have used the *RCL command to recall a state register that
wasn’t previously saved with the *SAV command.
–250
Mass storage error; Lock
violation
The process cannot access the file because another process has
locked a portion of the file.
–250
Mass storage error; Mkr
Table must be on to save
Mkr Table as Meas Results
You have to have a Marker Table on the screen before you can
save it. Turn on the Marker Table and try again.
–250
Mass storage error; No
file names available
Attempt to use the auto file name generation when all 10,000 file
names are taken.
–250
Mass storage error; Open
failed
The system cannot open the device or file specified. This could be
because the storage media is full, or possibly due to a filename
error. If using an external storage devise, check that the devise is
properly formatted.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...