114 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–224
Illegal parameter value;
Invalid list length
You are trying to set some list measurement settings, but the
multiple lists that you sent were not all the same length. The
number of settings must be consistent from list to list.
–224
Illegal parameter value;
LXI Event <event> already
exists.
This error occurs when you try to add an LXI Event that has
already been added.
–224
Illegal parameter value;
LXI Event <event> contains
illegal characters.
When a new LXI Event is created, it may not use the comma,
semicolon, or newline characters. All other printable ASCII
characters are valid.
–224
Illegal parameter value;
LXI Event <event> does not
exist.
The requested event has not been added yet.
–224
Illegal parameter value;
Measurement not available
You tried to turn on a measurement that is not available in the
current mode.
–224
Illegal parameter
value;This instrument is
always DC coupled
You can’t set AC coupling in this instrument
–224
Illegal parameter
value;This model is always
AC coupled
You can’t set DC coupling in this analyzer
–225
Out of memory
There is not enough memory to perform the requested operation.
–225
Out of memory;
Insufficient resources
to load Mode <mode name>
If you attempt to load a mode via SCPI that will exceed memory
capacity, the Mode does not load and this message is returned.
“mode name” is the SCPI parameter for the Mode in question, for
example, SA for Spectrum Analyzer Mode. You can free up
resources in the System, Power On, Configure Applications menu
–225
Out of memory; Memory
limit caused Data
Acquisition to be
truncated
–226
List not same length
You are using the LIST structure, but have individual lists that are
not the same lengths.
–230
Data corrupt or stale;
A legal data element was found, but it could not be used because
the data format or the data structure was not correct. Maybe a
new measurement had been started but had not completed.
–230
Data corrupt or stale;
Measurement data is not
available
Measurement data not available. The measurement that you are
trying to get data from must be the current active measurement.
Maybe you have not initiated the measurement, or it has not
completed all the sweeps/averages needed.
–230
Data corrupt or stale;
Trace contains no data.
Trace cannot be displayed because currently there is no data
assigned to it. Use the functions under the Trace menu, or load a
previously saved trace, to assign data to the trace.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...