556 N9030B PXA Signal Analyzer Service Guide
Post-Repair Procedures
Post-Repair Procedures
SW1
Preselector bypass
switch (Option
MPB)
Frequency Response (all tests)
Reset Low Noise Path switch using software
utility.
4800 MHz Calibrator Adjustment
Frequency Response (all tests)
SW2
Preselector bypass
switch (Option
MPB)
Frequency Response (all tests)
Reset Preselector Bypass cycles using
software utility.
4800 MHz Calibrator Adjustment
Frequency Response (all tests)
SW3
Low Noise Path
Switch (Option
LNP)
Frequency Response (all tests)
Reset Low Noise Path switch using software
utility.
Frequency Response (all tests)
a. Calibration Application Software is required for all performance tests and most adjustments. For
details go to
http://www.keysight.com/find/calibrationsoftware
b. See the Additional Tasks section in this chapter for detailed instructions on this task.
c. To set the instrument time and date press
Start
,
Control Panel
,
Date Time Language and
Regional Options
,
Date and Time
d. Before replacing the A5 Disk Drive, see the “Calibration Data Backup and Restore” section in this
chapter for details.
e. See Chapter 17, “Instrument Software” for instructions on updating the instrument software to the
latest version.
f. The 50 MHz and 4800 MHz Calibrator Amplitude Adjustments must be done prior to Frequency
Response Adjustments.
Some Adjustment tests may be found in the Utility portion of the Test
Plan.
Table 17-1
Post-Repair Testing Requirements (Continued)
Assembly
Adjustments to Perform
(in the order listed)
a
Performance Verifications to Perform After
Adjustments
a
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...