330 N9030B PXA Signal Analyzer Service Guide
Hardware Options
A25 Wideband Analog IF Troubleshooting - Option B2X and B5X
A25 Wideband Analog IF Troubleshooting - Option B2X and B5X
A25 Wideband Analog IF Assembly Description - Option B2X and B5X
The analyzer's RF input signal is down converted to a 250 MHz, 750 MHz, or
877.148 MHz intermediate frequency by the A13 Front End assembly and
routed through the A15 Front End controller to A15J901. Cable W70 connects
to A25J17 Wideband Analog IF Input. The signal is filtered, attenuation and
gain applied based upon the instrument settings. This signal is then sampled
by the ADC and the resultant data is the sent to the A26 Wideband Digital IF
(Option B2X) for processing via the flex circuit cable. If the instrument has
Option B5X, 510 MHz Analysis BW, the resultant data is sent to the A26 and
A27 Wideband Digital IF boards.The A26 and A27 Wideband Digital IF have no
service accessible locations and should be troubleshot from the front panel
controls.
If the analyzer is using the Option B40, 40 MHz IF Path, the signal at 250 MHz
IF frequency is switched to J18 and this 250 MHz IF is routed through W71 to
the A3 Digital IF assembly.
If the analyzer is using the Option B2X, 255 MHz IF Path, the signal at J17 will
be at 750 MHz. If the analyzer is using the Option B5X, 510 MHz Path, the
signal at J17 will be 877.148 MHz. Both of these IF signal frequencies are
processed through the A25 Wideband Analog IF assembly.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...