KMS-30/400 Supervisor's Manual
5-9
7.
Use the arrows on either side of the number read-out bar to adjust the offset
up or down so that the lower slide bar on the measurement screen is floating
just above the point at which it is no longer red (Figure 5-6).
Once adjusted correctly, the bar should be white and moving up and down
slightly.
Figure 5-6. Offset Adjust Value
Before
After
L
Note: Perform Step 8 only if you are configuring transmitted light.
8.
Ensure that the system has illumination values as shown in Table 5-1.
Table 5-1. Illumination Ranges
Objective Illumination
Gain
Offset
Analog
Gain
Camera
Gain*
Camera
Shutter*
10X 30-35 31-35
47-56 1 5-7 5-6
50X 45-55
31-35
47-56 1 10-15 6-7
100X
50-75
31-35
47-56 1 15-20 8
150X
60-90
31-35
47-56 1 15-20 8
* KMS-400 systems only
9.
Adjust the other objectives while the Illumination pop-up menu is still on
the screen, then rotate the turret to the next objective.
10.
Repeat steps 5-7 for each objective.
11.
To save illumination settings for the KMS-310:
a.
Press [ESCAPE] to exit the Illumination submenu.
b.
Press [ESCAPE] to exit the Measurement screen and return to the
Master Menu.
c.
Select
Save Program
and press [ENTER].
12.
To save illumination settings for the KMS-400 system:
Summary of Contents for KMS-310
Page 10: ...Contents viii...
Page 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Page 51: ...User Interface Overview Software Controls Software Organization 4...
Page 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Page 266: ...7 48 Chapter 7 Operation...
Page 292: ...8 26 Chapter 8 Maintenance...
Page 293: ...Error Messages System Error Messages Script Error Messages 9...
Page 297: ...Glossary...
Page 304: ...I 4 Index...