KMS-30/400 Supervisor's Manual
5-17
Configure Single-point Calibration
This section provides instructions for configuring single-point calibration and
includes the following procedures:
Set the illumination
Disable submicron multipoint calibration
Set up the measurement window
Calibrate the system to standard
Set the Illumination
To set illumination and check single-point mode status:
1.
Place the measurement gauge (or standard) on the stage.
2.
Focus the condenser assembly for the thickness of the standard.
3.
Set the illumination with the 100x objective.
Disable Submicron Multipoint Calibration
To disable submicron multipoint calibration:
1.
With the Measure screen displayed, press [ALT + O].
2.
View the comment line.
L
Note: If SUBM is not displayed in the comment line, then this option is not installed and
you can skip the rest of this procedure and go on to the next section.
If SUBM is displayed in the comment line, complete the remainder of this procedure to
disable the submicron multipoint calibration function.
3.
From the Supervisor Master Menu screen, highlight the
Program Setup
icon
and press [SELECT].
4.
With the Program Setup screen displayed, delete the file name displayed in
the
Multipoint File Name
.
5. Highlight
the
Exit
button and press [SELECT] to return to Supervisor
Master Menu screen.
Once the submicron multipoint calibration function has been disabled,
“SUBM FILE INVALID OR DOES NOT EXIST”
is displayed across
the right portion of the screen.
Summary of Contents for KMS-310
Page 10: ...Contents viii...
Page 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Page 51: ...User Interface Overview Software Controls Software Organization 4...
Page 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Page 266: ...7 48 Chapter 7 Operation...
Page 292: ...8 26 Chapter 8 Maintenance...
Page 293: ...Error Messages System Error Messages Script Error Messages 9...
Page 297: ...Glossary...
Page 304: ...I 4 Index...