KMS-310/400 Supervisor's Manual
6-93
SEQEDGE
FUNCTION
SEQEDGE: Vseq Vwin Vaxis Vdir Vedge Vth
Vseq Sequence number, 0-9
Vwin Measure window: 0 for top or left 1 for bottom or
right
Vaxis Measure 0 for horizontal, 1 for vertical
Vdir 0 for negative direction, 1 for positive
Vedge Edge number from neg/pos side of window, 1-100
Vth
Threshold percentage for edge detection
USE
50% threshold assumed unless overridden using Ctrl-t.
EXAMPLE SCRIPT
Figure 6-8 is an example of a sequence edge number script.
RELATED SCRIPTS
SEQWIN; RECALL
Figure 6-8. Example of a Sequence Edge Number Script
Edge 3
from Positive 1
start of Window 0
Window 0
Horizontal measure
SEQEDGE: 7 0 0 1 3 30
Summary of Contents for KMS-310
Page 10: ...Contents viii...
Page 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Page 51: ...User Interface Overview Software Controls Software Organization 4...
Page 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Page 266: ...7 48 Chapter 7 Operation...
Page 292: ...8 26 Chapter 8 Maintenance...
Page 293: ...Error Messages System Error Messages Script Error Messages 9...
Page 297: ...Glossary...
Page 304: ...I 4 Index...