KMS-30/400 Supervisor's Manual
5-31
Set Piezo Autofocus Algorithm
To set the Piezo using the autofocus algorithm (Figure 5-23):
Figure 5-23. Piezo Autofocus Algorithm
2*step
swing
range/2
Starting
position
Pass 3
Pass 1
Pass 2
stepsize = 110 nm
Final
position
stepsize = 'step'
stepsize = 7.324 nm
1.
Complete a first pass (pass 1).
Scan from start range/2 through start+ range/2 for the best focus position
(p1). To do this, take a focus reading approximately every 110 nanometers
(if you are using a 30 micron Piezo).
Use Table 5-3 and the following definitions to complete the first pass (p1):
Start:
initial position at the start of the autofocus procedure
Range:
a user-specified parameter
p1
: the best focus position
Table 5-3. Piezo Autofocus Algorithm
User Parameters
20 micron, 10 bit
30 micron, 12 bit
Min (nm)
Max (nm)
Min (nm)
Max (nm)
Range
: total scan distance of
pass 1
1000 20000 10
∅
0
30000
Swing
: half the total scan
distance of pass 2
100 15000 10
∅
15000
Step
: step size of scan during
pass 2
10 10000
7 15000
Summary of Contents for KMS-310
Page 10: ...Contents viii...
Page 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Page 51: ...User Interface Overview Software Controls Software Organization 4...
Page 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Page 266: ...7 48 Chapter 7 Operation...
Page 292: ...8 26 Chapter 8 Maintenance...
Page 293: ...Error Messages System Error Messages Script Error Messages 9...
Page 297: ...Glossary...
Page 304: ...I 4 Index...