KMS-30/400 Supervisor's Manual
5-21
Calibrate the System to Standard
To calibrate the system to standard:
1.
From the Calibration screen, highlight the
Calibrate
button and press
[SELECT].
The Calibration Control submenu is displayed.
2.
Highlight the
icon and recall Window 1.
3.
Highlight the
icon and move the Measure window as shown in Figure
5-16. Make sure that the correct edges are for pitch.
Figure 5-16. Measure Window Movement
Line
Pitch
4.
Set the scan direction for measurement using the standards listed in
Table 5-2.
Table 5-2. Scan for Measurement Standard Directions
Standard
Left
Right
RKS
Æ
Æ
NBS
Æ
Å
NIST
Æ
Å
5.
From the Measure screen, highlight the
icon and press [SELECT].
The following message will be displayed in the comment line once pitch
calibration is completed :
HORIZONTAL PITCH = =0.0567 MICORNS/PIXEL