5-32
Chapter 5: System Setup
2.
Complete a second pass (pass 2).
Scan from p1-swing through p1_swing for the best focus position (p2). To
do this, take a focus reading at every user-specified step.
Use Table 5-3 and the following definitions to complete the first pass (p2):
Swing
: initial position at start of the autofocus procedure
Step
: user-specified parameter
p2
: best focus position
3.
Complete a third pass (pass 3).
Scan from p2-(2*step) through p2+(2*step)for the best focus position (p3).
To do this, take a focus reading every 7.324 nanometers.
4.
Position to p3
.