KMS-310/400 Supervisor's Manual
6-97
STATMSR3
FUNCTION
Get statistics for a sequence window/collector
STATMSR3: Vseq[0-9] Vmax Vmin Vavg Vcnt Vsdev
Vseq Sequence window number/collector
Vmax Maximum measured value in the collector
Vmin Minimum measured value in the collector
Vavg Average value of all the measurements
Vcnt Number of measurements collected
Vsdev 3 sigma standard deviation
USE
Retrieves the statistics for the specified collector. Returns
maximum measurement, minimum measurement, average
measurement, number of measurements, and a 3 sigma value for
standard deviation. The lsd of the standard deviation for
STATMSR3 is one thousandth of a nanometer. STATMSR3 and
STATMSR differ only in the standard deviation returned.
STATMSR returns one sigma and the lsd is one nanometer.
OPTIONS
N/A
EXAMPLE SCRIPT
MATH: V3298=2 /select
collector
#2
SEQWIN: V3298 1 100 100 200 200
SEQWIN: V3298 0 125 125 175 175
MEASURE: 3
STATMSR3 V3298 V12 V13 V14 V15 V16
MESSAGE: “3 SIGMA STD DEV = %”, V16
:::
EXIT:
RELATED SCRIPT
STATMSR
Summary of Contents for KMS-310
Page 10: ...Contents viii...
Page 33: ...System Overview Product Overview Subsystem Overview Functional Overview 3...
Page 51: ...User Interface Overview Software Controls Software Organization 4...
Page 119: ...Creating Automated Scripts Overview Script Creation Script Locator Script Commands 6...
Page 266: ...7 48 Chapter 7 Operation...
Page 292: ...8 26 Chapter 8 Maintenance...
Page 293: ...Error Messages System Error Messages Script Error Messages 9...
Page 297: ...Glossary...
Page 304: ...I 4 Index...