3.15.3. Unit Dismantling and Re-assembly
1.
Unit
dismantling
1. Turn off system power switch and PC
power
switch.
2. Turn off the system main power.
3. Remove the lower rear panel.
4.
Loosen screw fixing DP-1, and take
it off, and pull out DP-1 a little
towards
you.
5. Remove piping (four) fixed to unit.
When disconnecting tube, waste
water in the tube flows out.
Wipe it off with tissue paper.
6.
Remove connector fixed to DP-1.
J5501
7.
Pull out DP-1 towards you to remove
it.
8.
The explanation for removal of DP-2
is omitted.
The procedure is the same as of DP-1.
Note Take a great care when removing
because waste water might cause
infection.
Fig. 3-90 DP-2
Joint of Piping
Fixing Screw
Connector
Fig. 3-89 DP-1
Fixing Screw
Fig. 3-88 System back view
DP-1
DP-2
Joint of Piping