3‑36
3.8.3. Unit Dismantling and Re-assembly
1. Dismantling
1. Loosen fixing screws (six) of plate and
remove the screws.
2.
Remove
the
plate.
3.
Loosen fixing nuts (two) of mixing
base of MU-1, and remove them.
And take out mixing base with
piping.
4.
Loosen fixing nuts (two) of mixing
base of MU-2, and remove them.
And take out mixing base with
piping.
Fig. 3-54 MU-1
Fixing Nuts
Mixing Base
Fig. 3-55 MU-2
Mixing Base
Fixing Nut
Fig. 3-53 Reaction Tray periphery
Fixing Screws (six)
Plate