Service Manual
Chapter 3
3‑21
9.
The left side view shows reagent
tray without Peltier element
10. Apply heat sinker to both sides of
new Peltier element thinly and
uniformly.
11.
Set Peltier element as shown on the
left side, while holding down.
Be sure to put Peltier element
with the letters on it up.
12.
Assemble with the reverse procedure
of the above.
When screwing M3 screws (three),
use torque driver (4 kg) to tighten
them
uniformly.
Fig. 3-30
Fig. 3-31
Connector
Setting position of Peltier element
Fig. 3-29
The side with letters
is shown up.