Service Manual
Chapter 3
3‑3
1. Sample trays
Patient tray (In case of No. 1 tray)
(Outside) Patient sample (1 - 30) -------------30 samples
(Inside)
Patient sample (31 - 40) -----------10 samples
Control sample (C7 - C12) ---------- 6 samples
Washing (W1, W2) -------------2 samples
STAT (E1 - E5) ------------- 5 samples
ISE standard liq. (CAL) ------------ 1 sample
ISE washing liq. (WASH) ---------- 1 sample
Patient tray (In case of No. 2 tray)
(Outside) Patient sample (41 - 70) ---------- 30 samples
(Inside)
Patient sample (71 - 80) -----------10 samples
Control sample (C13 - C18) --------- 6 samples
Reserved (W1, W2) ------------2 samples
STAT (E6 - E10) ----------- 5 samples
ISE washing liq. (WASH) ---------- 1 sample
2 patient trays (No. 1 and No. 2) are attached as standard accessories, and
other 8 trays (No. 3 - No. 10) can be supplied as option.
Calibration
tray
(Outside) Blank (B1, B2)-----------------------------2 samples
Standard sample (S1 - S28) ------- 28 samples
(Inside) Standard sample (S29 - S45) -------- 17 samples
Control sample (C1 - C6) ------------6 samples
ISE standard liq. (CAL) ------------1 sample
2. Sample tray discrimination
Sample tray has combination slit which corresponds to tray number.
When a tray is set to a sampler, a sensor on the STF PCB
detects this slit and recognizes tray kind and number.
Summary of Contents for Labmax 240
Page 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Page 12: ......
Page 18: ......
Page 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Page 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Page 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Page 46: ......
Page 110: ......
Page 140: ......