5‑14
5.6.2 STM-R (Sample Probe - Rotation)
1. Explanation of Function
2. INIT
To move to the home position. (And to stop when detecting the
home sensor.)
3. OFFSET
To move for the corrective value set in the offset screen.
4. ADJUST
To input the angle to move by value, and to select the direction
of rotation.
5. D-CELL -> CELL To move from the reaction cuvette (dilution position) to the
reaction
cuvette.
6. WASH -> OUT-S
To move from the washing trough to the outside sample cup.
7. WASH -> IN-S
To move from the washing trough to the inside sample cup.
8. OUT-S -> CELL
To move from the outside sample cup to the reaction cuvette.
9. IN-S -> CELL
To move from the inside sample cup to the reaction cuvette.
10. CELL -> WASH
To move from the reaction cuvette to the washing trough.
11. OUT-S -> ISE
To move from the outside sample cup to ISE.
12. IN-S -> ISE
To move from the inside sample cup to ISE.
13. ISE -> WASH
To move from ISE to the washing trough.
14. OUT-S -> DCELL To move from the outside sample cup to the reaction cuvette
(dilution
position).
15. IN-S -> DCELL
To move from the inside sample cup to the reaction cuvette
(dilution
position).
16. CELL -> IN-S
To move from the reaction cuvette to the inside sample cup.
17. CELL -> OUT-S
To move from the reaction cuvette to the outside sample cup.
18. Low Speed
* Starting pulse rate
19. High Speed
* The highest speed pulse rate
20. Pulse Count
* Adjustable speed pulse count
21. Init Speed
* Pulse rate of initialization
* shows that setting is not needed.
Manual Function (STM-R)
Summary of Contents for Labmax 240
Page 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Page 12: ......
Page 18: ......
Page 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Page 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Page 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Page 46: ......
Page 110: ......
Page 140: ......