Service Manual
Chapter 3
3‑37
5.
Loosen fixing screws (two) of MU-1
and remove them.
And then remove unit
6.
To remove MU-2, take off front panel
first.
And loosen fixing screws of upper
panel to remove them, and take off
upper
panel.
Refer to Chapter 1, page 1-5.
7.
After loosening fixing screws (two) of
MU-2, take them off.
And then remove MU-2.
2.
Assembling
Assembling procedure is the reverse of the above.
3.
Cautions
Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
Fig. 56 MU-1
Fixing Screw
Fixing Screws
Fig. 58 MU-2
Fig. 57 System slant view