3‑30
Cuvette Holder
Motor
Reaction Cuvette
3.7. Reaction Tray (RCT)
3.7.1. Parts and Configuration
- Refer to “ASSY-9-1/2, 2/2 RCT” for the unit assembly drawing.
- Refer to “WIRING-12 RCT” for the unit wiring diagram.
- Refer to “Operation manual 5.2.3 Reaction Cuvette” for the cuvette exchange.
3.7.2. Function and Action
1. Function
This tray can hold sixty (60) reaction cuvettes and keep the temperature
in cuvettes at 37 Celsius degrees.
In these cuvettes, reagent and samples are made react.
Reaction cuvette is optical cell for absorbance measurement.
(Reaction cuvette direct photometry method)
Fig. 3-42 Reaction Tray
Fig. 3-41 Reaction Tray periphery