0‑8
6. Installation Requirements
Only the trained and authorized serviceman is permitted to install or move
the system, following the requirements below.
3.1 Power supply requirements
(1) Power supply capacity should be over 600VA.
(2) Voltage : AC230V plus-minus 10%
Frequency : 50/60Hz plus-minus 1Hz.
(3) Earth resistance of ground terminal should be less than 10 ohm.
(4) Power must be supplied for 24 hours a day.
(5)
Install away from high frequency electromagnetic wave emitting equipment
(such as centrifuge, electric discharge), preferably not in the same room.
3.2 Required conditions
(1)
The system should be installed at altitude less than 2000 meters.
(2)
The system can be put in operation between 5 and 40 Celsius degrees.
But its performance is guaranteed only when room temperature is 15 ~ 30
Celsius degrees and the temperature variation during the operation should be
less than plus-minus 2 Celsius degrees/H.
(3) Room humidity should be 40 ~ 80 %, and no condensation.
(4) The table should be able to support over 100kgf/m
2.
(5) The gradient of the table should be less than 5mm/m.
(6)
Keep the space around the system for taking out the power cord
in an emergency, for good ventilation and for maintenance, as the following.
(7)
The following points should be taken into consideration to select the
installation
position.
- No dust, good ventilation
- No direct sun light
-
Small
vibration
3.3 Water supply/drain requirements
Water reservoir and waste liquid reservoir should be set below the system
bottom
level.
32 cm
100 cm
50 cm
ANALYZER
FRONT
AC INPUT
50 cm
Also keep the space of at
least 100 cm over the
system.
(At the minimum)
Summary of Contents for Labmax 240
Page 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Page 12: ......
Page 18: ......
Page 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Page 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Page 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Page 46: ......
Page 110: ......
Page 140: ......