5‑18
5.6.6 RTM - R (Reagent Probe - Rotation)
1. Explanation of Function
2. INIT
To move to the home position. (And to stop when detecting the
home
sensor.)
3. OFFSET
To move for the corrective value set in the offset screen.
4. ADJUST
To input the angle to move by value, and select the direction of
rotation.
5. CELL -> WASH
To move R1 probe from the reaction cuvette to the washing
trough.
6. WASH -> BOTL
To move R1 probe from the washing trough to the bottle 1 (R1).
7. BOTL -> CELL
To move R1 probe from the bottle 1 to the reaction cuvette.
8. WASH -> IDLE3
To move R2 probe to R2 bottle position.
9. IDLE3 -> IDLE1
To move R2 probe to the reaction cuvette.
10. IDLE2 -> BOTL
To move R1 probe to the bottle 1 position.
11. IDLE1 -> IDLE2
To move R1 probe to the washing trough.
12. Low Speed
* Starting pulse rate
13. High Speed
* The highest speed pulse rate
14. Pulse Count
* Adjustable speed pulse count
15. Init Speed
* Pulse rate of initialization
* shows that setting is needed.
Manual Function (RTM – R)