5‑12
5.6 Outline of Manual Operation Screen
1.
When clicking M.Function button in the enlarged sub-menu of Maintenance,
the following screen is shown.
The basic procedure of manual function is as follows.
Unit
→
Function
→
RUN
Unit (abbreviations) (units) (movement)
1. STF
Sampler
Rotation of sampler
2. STM-R
STM Rotation
Rotation of sample probe
3. STM-UD
STM up/down
Up/Down movement of sample probe
4. SP
Sample pump
Aspiration/Dispensation movement of sample pump
5. RGT
Reagent tray
Rotation of reagent tray
6. RTM-R
RTM rotation
Rotation of reagent probe
7. RTM-UD
RTM up/down
Up/Down movement of reagent probe
8. SP
Reagent pump
Aspiration/Dispensation movement of reagent pump
9. RCT
Reaction tray
Rotation of reaction tray
10. MU-1
Mixing unit-1
Up/Down movement of mixing unit-1
11. MU-2
Mixing unit-2
Up/Down movement of mixing unit-2
12. MP
Mixing pump
Movement of mixing pump
13. CWS
Cuvette washing station Up/Down movement of reaction cuvette
washing
station
14. CWP
Cuvette washing pump
Aspiration/Dispensation movement of reaction
cuvette washing pump
15. PWP
Probe washing pump
Aspiration/Dispensation movement of probe
washing
pump
16. DP 1/2
Drainage pump-1, 2
Movement of drainage pump-1 and 2
17. Detector
Colorimetry detector
Not ready.
18.
TEST
?
Not
ready.
19. EVS
Electromagnetic valve
Not ready
monitor
Manual Function screen
Sensor display
When it is ON,
the letters are
shown in red.
1. Selection
of unit.
2. Selection
of function.
3. Perform
function
.
For initialization
of all the units.
Click, when
operating with
“NO Safety”.
Sub-screens
Summary of Contents for Labmax 240
Page 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Page 12: ......
Page 18: ......
Page 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Page 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Page 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Page 46: ......
Page 110: ......
Page 140: ......