3‑54
The temperature is set a little higher than 37 Celsius degrees so as to keep
washing water at 37 Celsius degrees when dispensing to reaction cuvettes.
As protection function, thermostat (10) is installed.
In case the temperature exceeds 70 Celsius degrees, power source will be turned
OFF not to electrify to heater (4).
Moreover, to prevent heating empty lines, level sensor (12) is installed in order to
check whether there is any water in it or not