5‑20
5.6.8 RP (Reagent Pump)
1. Explanation of Function
2. INIT
The washing side of pump : To move to the home position
(the upper limit).
3. PULL(1800uL)
The washing side of pump : Aspiration movement (1800uL)
4. PUSH(1800uL)
The washing side of pump : Dispensation movement (1800uL)
5. Move Speed
The washing side of pump : * Move pulse rate
6. INIT
The reagent side of pump : To move to the home position
(the upper limit).
7. PULL(75uL)
The reagent side of pump : Aspiration movement (75uL)
8. PUSH(75uL)
The reagent side of pump : Dispensation movement (75uL)
9. Volume Set
The reagent side of pump : Setting of the amount of reagent for
aspiration or dispensation.
CW means Dispensation and CCW means Aspiration.
10. Move Speed
The reagent side of pump : * Move speed pulse rate
* shows that setting is not needed.
11. EV2-ON
Electromagnetic valve EV-2 ON
Line switch to the probe.
12. EV2-OFF
Electromagnetic valve EV-2 OFF
Line switch to the
water
tank.
13. EV3-ON
Electromagnetic valve EV-3 ON
Line switch to the R2
probe.
14. EV3-OFF
Electromagnetic valve EV-3 OFF
Line switch to the R1
probe.
Manual Function (RP)