4‑26
3. When breaking down of this board
When this PCB is broken down, one or more of the following phenomena
may
occur.
1. Reaction cuvette washing pump (CWP) won’t function, or won’t stop functioning.
- Washing solution does not come out from dispensation nozzle of reaction
cuvette washing station (CWS), or is flowing all the time.
2. Waste liquid pump won’t function.
- Waste liquid is overflowed from probe washing trough.
3. The temperature of washing solution does not rise to the set temperature,
or rises abnormally high.