Service Manual
Chapter 3
3‑17
3.4.3. Unit Dismantling and Re-assembly
1. Unit dismantling
1.
Turn off system power source and PC
power switch.
2.
Open top cover and push over upper
chassis gently to give a tilt.
- Make sure to remove fixing screws on
left and right panels in advance.
- Confirm the “click” sound made by a
stay when fixed.
3. Remove front panel first, and then
upper panel.
-
Refer to Chapter 1, page 1-5.
4. Hold bottle case by edge to lift up and
remove bottle case.
5. Remove the following wiring
connectors.
J4706 J4707 J4701 J4702
J4703
6.
Loosen unit fixing screws (four), and
remove them.
7. Hold the upper part of outer case, and
lift up unit to remove.
2. Assembling
Assembling procedure is the reverse of the above.
Bottle Case
Outer Case
Fig. 3-22 Reagent Tray periphery
Connector
Fixing Screws (four)
Fig. 3-21 System front view
Summary of Contents for Labmax 240
Page 1: ...Automated Clinical Analyzer Service Manual Version 1 44 TOKYO BOEKI LTD...
Page 12: ......
Page 18: ......
Page 40: ...2 8 2 4 Reagent Dispensation Process Fig 2 4 Reagent Dispensation Process 1 2...
Page 41: ...Service Manual Chapter 2 2 9 Fig 2 5 Reagent Dispensation Process 2 2...
Page 44: ...2 12 Fig 2 7 Sample Dispensation Process 2 2...
Page 46: ......
Page 110: ......
Page 140: ......