3‑20
5.
Take out the plate and then insulator.
6. Take out revolving shaft.
Leave temperature sensor set in
revolving shaft as it is, and put this
revolving shaft nearby.
7.
Take out connector of Peltier
element.
8. Take out Peltier element.
Fig. 3-26
Insulator
Revolving shaft
Fig. 3-27
Revolving Shaft
Fig. 3-28
Connector