Service Manual
Chapter 4
4‑29
4.10 Shift of AC Input Voltage
Three kinds of voltage, AC230V, AC115V and AC100V are available
to this system.
Rear side view
Rear side view
Signal wire connection
Power Supply Panel (Left rear)
RS-232C cable
Main Power Switch
(Shows OFF state.)
AC Input
Fuse
T-250V 6.3A.
AC230V, AC115V and AC100V
terminals are here.
Available voltage is factory set
according to your country use
is shown on the label.