Service Manual
Chapter 3
3‑33
8.
After removing fixing nuts (two) of
MU-1, remove MU-1.
- Lay removed MU-1 nearby lightly
with mixing base and piping.
9.
After removing fixing nuts (two) of
MU-2, remove MU-2
- Lay removed MU-2 nearby lightly
with mixing base and piping.
10.
Remove fixing screws (four) of
joining
plate.
Fig. 3-46 MU-1
Fixing Nut
Fig. 3-47 MU-2
Fixing Nut
Mixing Base
Fig. 3-48 Lamp House periphery
Fixing Screws (four)
Joining Plate
Lamp House