Service Manual
Chapter 5
5‑13
5.6.1 STF (Sampler)
1. Explanation of function
2. INIT
To move to the home position. (And to stop when detecting the
home sensor.)
3. OFFSET
To move for the corrective value set in the offset screen.
4. ADJUST
To input the angle to move by value, and to select the direction
of rotation.
5. CW1Pitch(OUT)
To move the outside cups clockwise by one pitch.
6. CW1Pitch(IN)
To move the inside cups clockwise by one pitch.
7. CCW1Pitch(OUT) To move the outside cups counter-clockwise by one pitch.
8. CCW1Pitch(IN)
To move the inside cups counter-clockwise by one pitch.
9. POSITION
10. Low Speed
* Starting pulse rate
11. High Speed
* The highest speed pulse rate
12. Pulse Count
* Adjustable speed pulse count
13. Init Speed
* Pulse rate of initialization
* shows that setting is not needed.
Manual Function (STF)