3‑48
5. Remove connectors (two).
J4801 J4803
6. Loosen unit fixing screws (three) and
remove
them.
7. Lift up unit and remove it.
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
Fig. 3-72 CWS Base Plate
Connector Fixing
Screw