5‑30
5.7 Manual Function Sub-screens
5.7.1 Encoder Screen
1.
Explanation of the screen
When clicking Encoder button in the bottom of the manual function screen,
the following screen appears.
This screen is used for checking of encoder.
Encoder screen
1. Offset
Not ready. (0 is shown.)
2. Theory
The theoretical value is shown.
3. Actual measurement
The actual measurement value is shown.
4. Error Limit
The limit is shown. If the measurement exceeds this
value, it causes an error.
5. Error with the theory
The sum of the theoretical value and the limit value
is shown.
6. Position
The position number at the time of clicking is shown.
7. Read
The values of the above mentioned 1 – 6 at the time of
clicking are shown.
8. Exit
When clicking, the screen returns back to the manual
function screen.
Encoder screen
Unit names
The values
are shown.